Beckman Coulter AU5831 Clinical Chemistry Analyzer

Beckman Coulter AU5831 Clinical Chemistry Analyzer
Contact us for price
Description
Includes:
– One Beckman Coulter AU5831 with Single ISE, Instrument and Peripherals, 210V
The fastest full-menu clinical chemistry analyzers.
Give your high- and ultra high-volume core laboratory a performance boost. Clinical chemistry analyzers in the AU5800 series are the fastest in the industry, processing up to 2,000 tests per hour with a full menu of assays, including Therapeutic Drug Monitoring (TDM) and Drugs of Abuse. Increase lab productivity with continuous operation, powerful throughput, and scalable capacity up to four modules.
Features
- Ideal For: Hospital or reference laboratories (4.0M to 6.5M tests/year)
- Number of Onboard Reagents: 162
- Photometric Throughput: Up to 6,000 tests/hour
- Throughput with Dual ISE: Up to 7,800 tests/hour
- Number of Simultaneous Assays: 120
- Sample Capacity: 400 samples (2×20 rack set)
Test Menu
Specifications: System Specifications Reaction System Specifications Analytical Method Specifications Type of Measurement: End point assay Rate assay Fixed point assay Electrode method (ISE) (option) Optical System Specifications
Specifications
| Condition | New |
| Stock Number | 247 |
| Reaction Incubation Method | Dry bath system |
| Reaction Temperature | Dry bath: 37± 0.3 °C (98.6± 0.5 °F) |
| Reaction Solution Amount | • 80 to 287 µL |
| Reaction Time | Maximum 8 minutes 40 seconds |
| Reaction Cell | Glass square cuvette (optical path length: 5 mm) |
| Reaction Line | Rotary disk system: 204 cuvettes x 2 lines |
| Mixing System | Rotative mixing bar system |
| Photometer | Multi-wavelength diffraction grating spectrophotometer Photometric Modes: Monochromatic or bichromatic Wavelengths 340 to 800 nm (13 steps of 340, 380, 410, 450, 480, 520, 540, 570, 600, 660, 700, 750, and 800 nm) |
| Measurable Absorbance Range | 0 to 3.0 Abs (converted in units of 10 mm of optical path length) |
| Photometric points | 28 |
| Photodetector | Silicon photodiode array |
| Light Source | Halogen lamp 12 V/100 W |
| Photometric resolution | 0.0001 Abs |