Horiba ABX Pentra 60C+ Hematology Analyzer

Horiba ABX Pentra 60C+ Hematology Analyzer
Contact us for price
or
Call 951.296.1266
Description
Horiba ABX Pentra 60C+ Hematology Analyzer
Includes:
- One Pentra 60C+ complete with accessories
- A Dedicated Workstation for a Powerful Management of your Workflo
- Closed tube
- 26 parameters
- CBC / 5 Part DIFF
- Data management on stand-alone PC.
- Micro-sampling from whole blood (CBC : 30 µL – DIFF: 53 µL)
- Cytochemistry, Impedance (real cell volume measurement) & Optical (analysis of the internal cellular structure by measuring light absorbance) – DHSS* Technology.
- Perfect homogenization of blood samples with reagents – MDSS** Technology.
DHSS : Double Hydrodynamic Sequential System
MDSS : Multi-Distribution Sampling System
Features/Benefits
- Throughput: Up to 60 samples/hour.
- Reagents: Only 4 onboard reagents and 1 diluent.
- Perfect differentiation of the 5 WBC sub-populations with DHSS*
- 3 histograms for RBC, BAS/WBC and PLT together with the 5 DIFF
- Basophils counted through specific channel.
- High resolution matrix.
Specifications:
Sample Specifications
- Throughput: 60 samples/hour (closed tube)
- Sample Mode: Closed tube
- Sample Identification: Alphanumeric; optional barcode reader
- Blood Volume:
- CBC mode: 30 µL
- CBC + DIFF mode: 53 µL
- Sample Stability: 48 hours post-draw stability
Parameters — 26
- WBC
- NEU# & NEU%
- LYM# & LYM%
- MON# & MON%
- EOS# & EOS%
- BAS# & BAS%
- ALY# & ALY%*
- LIC# & LIC%*
- RBC
- HGB
- HCT
- MCV
- MCH
- MCHC
- RDW
- PLT
- MPV
- PCT*
- PDW*
* RUO parameters (Research Use Only)
Graphs
- High-definition LMNE matrix
- PLT curve
- RBC curve
- WBC/BAS curves
Results & Performance
Parameter
Precision (%CV)
Linearity
WBC
< 2.0
0–120 × 10³/µL
RBC
< 2.0
0–8 × 10⁶/µL
HGB
< 1.0
0–24 g/dL
HCT
< 2.0
0–67%
PLT (whole blood)
< 5.0
0–1,900 × 10³/µL
PLT (PLT concentrate)
< 5.0
0–2,800 × 10³/µL
Reagents
- 1210906003 — ABX Basolyse
- 1210903010 — ABX Cleaner
- 1210901020 — ABX Diluent
- 1210206010 — ABX Eosinofix
- 1210906013 — ABX Lysebio
- 1210401005 — ABX Minoclair
Blood Controls & Calibrators
Blood Controls
- 5300000490 — ABX Difftrol Tri-Level, 3 × 3 mL vial
- 5300000496 — ABX Difftrol Tri-Level, 6 × 3 mL vial
- 5300000502 — ABX Difftrol Tri-Level, 12 × 3 mL vial
Calibrator
- 5300000276 — ABX Minocal Whole Blood Hematology Calibrator, 2 × 2.5 mL
Technical Specifications
- Weight: 77 lbs (35 kg)
- Dimensions: 20.3 × 17.5 × 19 in (51.5 × 44.5 × 48 cm)
- Power Supply: 100–240 V (±10%), 50–60 Hz
- Printer: Laser printer
- Throughput: 60 tests/hour
Data Management
- 15-inch color monitor
- Resolution: 800 × 600 px
- Windows XP Professional™
- User-defined flagging limits
- Patient & QC data transmission to LIS
- Uni-directional or bi-directional connection
- Communication protocols: ABX / ASTM
- CD-ROM drive / floppy disk
- Minimum 2 serial ports
- USB keyboard and USB mouse
Memory
- Stores up to 10,000 results with graphics
Quality Control Management
- 12 selectable QC files
- XB: 60 operator-selectable batches with statistics
- 20 samples per batch
- Within-run QC
- Levey-Jennings graphs
Logs
- Reagents
- Calibration
- Maintenance
- Errors
- Blank cycles
Specifications
| Manufacturer | Horiba |
| Model | Pentra 60C+ |
| Condition | Refurbished |
| Stock Number | 97 |
| Weight | 77lbs |
| Power Supply | 100 to 240 V (± 10%) - 50 to 60 Hz. |
| Dimensions | 20.3 x 17.5 x 19 in 51.5 cm x 44.5 cm x 48 cm |
| Printer | Laser |
| Throughput | 60 tests/hour |
| Memory | 10,000 results + Graphics |
| Logs | Reagents, calibration, maintenance, errors, blank cycle |
| DHSS | Double Hydrodynamic Sequential System |
| MDSS | Multi-Distribution Sampling System |
| Sample Mode | Closed tube |
| CBC mode | 30 µL |
| Sample Identification | Alphanumeric; optional barcode reader |
| Resolution | 800 × 600 px |
| Communication protocols | ABX / ASTM |
| Sample Stability | 48 hours post-draw stability |