Harvard Apparatus Pump 11 Plus Syringe Pump Single/Dual Infusion CE
Includes:
The Pump 11 Plus is a powerful yet easy-to-use syringe pump featuring a two-line, 16-character vacuum fluorescent display and a six-key membrane keypad. Operation is simple, requiring only syringe inner diameter and flow rate to begin pumping, with the ability to adjust flow rate while running. It offers two operating modes—Constant Flow Rate for continuous operation and Volume Dispense for precise fluid delivery. Micro-stepping pump profiles ensure smooth, consistent flow across the full range, and non-volatile memory retains the last syringe size, flow rate, and configuration settings.
The Pump 11 Plus is available in Standard and Advanced models, each offered in single or dual syringe configurations to suit various applications. The Standard model supports infusion only and includes an end-of-travel limit stop and anti-siphon bracket, while the Advanced model adds infusion/withdrawal capability and dual RS-232 serial ports. Designed with safety in mind, the motor stalls without damage during overload, and a manual adjustment collar prevents the syringe plunger from bottoming out. The Pump 11 Plus is CE Mark approved and meets all relevant European standards.
Features:
• Infusion only or infusion/withdrawal models
• Small footprint
• RS-232 option for computer control
• Drug/Nutritional studies
• Microfluidics/Nanofludics
• ESI Mass Spec
• Electrospinning
• MS Calibration
Specifications
| Manufacturer | Harvard Apparatus |
| Model | 11 Plus |
| Condition | Refurbished |
| Serial Number | B-59900 |
| Stock Number | 437 |
| type | Microprocessor-controlled syringe pump |
| Configuration | Single or dual syringe |
| Operation Modes | Infusion only or infusion/withdrawal |
| Accuracy | ±0.5% |
| Reproducibility | ±0.1% |
| Memory | Non-volatile memory (stores all settings) |
| Syringe Type | Plastic or glass |
| Single Syringe Size | 0.5 µL to 50/60 mL |
| Dual Syringe Size | 0.5 µL to 10 mL |
| Single Syringe | 0.0014 µL/hr to 26.56 mL/min |
| Dual Syringe | 0.0014 µL/hr to 7.91 mL/min |
| Average Linear Force | 16 lbs |
| Drive Motor | 0.9° step angle motor |
| Motor Drive Control | 1/4 micro stepping |
| Motor Steps per Leadscrew Revolution | 3,200 steps |
| Step Resolution | 0.33 µm/step |
| Input Power | 12 VDC, 1.5 Amps |
| Voltage Range | 100–240 VAC, 50/60 Hz (Universal Input) |
| Power Consumption | 18 watts |
| Power Supply | Harvard Apparatus–approved supply and line cord only |
| Dimensions (H × W × D) | 13 × 22.9 × 11.4 cm (5 × 9 × 4.5 in) |