Siemens Advia Centaur XPT Immunoassay System

Siemens Advia Centaur XPT Immunoassay System

Contact us for price

Description

ADVIA Centaur XPT Immunoassay System

Designed with one of the most advanced software packages on the market, the ADVIA Centaur® XPT System is among the highest-throughput systems available. The ADVIA Centaur XPT System delivers the results that clinicians depend upon for accurate diagnoses and better patient care—and does so predictably and consistently.
Siemens unites innovative workflow solutions with clinical excellence in the ADVIA Centaur XPT System, leading to greater laboratory productivity to stay ahead of increasing workload demands.

  • Simplified operation enhances lab efficiency.
  • Cutting-edge immunoassay testing supports clinicians in the screening, diagnosis, and monitoring of the most complex diseases.
  • Powerful automation and software streamline and accelerate your workflow.
  • Test Menu

    Anemia Special ID Cardiac
    Active B12 EBV-EBNA IgG** BNP
    Ferritin EBV-VCA IgG** CKMB
    Folate EBV-VCA IgM** High-Sensitivity Troponin I (TNIH)
    RBC Folate Syphilis Myoglobin
    Vitamin B12 Zika IgM†† TnI-Ultra™
    Hepatitis Oncology Thyroid
    Anti-HBs2 AFP Anti-TG
    HAV IgM BR 27.29 Anti-TPO
    HAV Total CA 125II‡ Free T3
    HBc IgM CA 15-3 Free T4
    HBc Total CA 19-9 T Uptake
    HBe Ag Calcitonin Total T3
    HBs Ag Confirmatory CEA Total T4
    HBs AgII Complexed PSA TSH3-Ultra
    HCV*** PSA TSH
    Serum HER-2/neu
    Reproductive Endocrinology Therapeutic Drug Monitoring ToRCH
    AFP Carbamazepine CMV IgG
    Anti-Müllerian Hormone** Digitoxin CMV IgM**
    DHEAS Digoxin Herpes-1 IgG
    Enhanced Estradiol Gentamicin Herpes-2 IgG
    Free βhCG‡‡ Phenobarbital Rubella IgG
    FSH Phenytoin Rubella IgM
    hCG Theophylline Toxoplasma IgG
    LH Valproic Acid Toxoplasma IgM
    PAPP-A‡‡ Vancomycin
    PlGF**
    Progesterone Immunosuppressant Drugs Bone Metabolism
    Prolactin Cyclosporine Intact PTH
    sFLT-1** Everolimus** Vitamin D Total
    SHBG Tacrolimus**
    Testosterone II
    Diabetes Metabolic Inflammation
    C-Peptide Cortisol IgE, Total
    Insulin Homocysteine
    HIV Autoimmune
    HIV 1/O/2 Enhanced (EHIV)§ Anti-CCP
    HIV Combo§

    Specifications

    ManufacturerSiemens
    ConditionNew
    System DescriptionRandom-access immunoassay system
    Test ThroughputUp to 240 tests per hour
    Time to First Result18 minutes; results every 15 seconds thereafter, assay dependent
    Assays Onboard30 assays
    Disease-State Assay GroupsAllergy, Anemia/Iron Metabolism, Bone Metabolism, Cardiac, Diabetes, Hepatitis, Immunosuppressant Drugs, Metabolic, Oncology, Reproductive Endocrinology, Special ID, Therapeutic Drug Monitoring (TDM), Thyroid, ToRCH
    Continuous OperationLoading/unloading of samples and consumables at any time without interrupting system operation. Includes sample racks, STATs, reagent packs, ancillary reagent packs, reaction cuvettes, sample pipette tips, calibrators, controls, wash fluids, water, and waste.
    Sample Tubes3 mL, 5 mL, 7 mL, and 10 mL tubes; 1 mL and 2 mL sample cups; microcontainer tubes
    Sample InputUniversal 5-position rack holds multiple tube types; 180 samples fully loaded; no-pause loading and unloading
    Validated Sample TypesSerum, plasma, urine (assay-dependent)
    Sample Volume per Test10–200 μL, assay-dependent
    Sample Integrity ControlPressure transducer technology with clot detection, flagging, and management; short-sample detection
    STAT HandlingDedicated STAT port accepts samples any time; STAT samples are processed with priority
    Auto-repeatUser-definable
    Sample DilutionAutomatic dilution varies by assay, up to 1:2500
    Auto-reflex TestingAutomatic ability to perform additional tests based on results of first test
    Sample Pipette Tips840 disposable sample tip capacity
    Sample Carryover PreventionDisposable pipette tips ensure zero sample carryover
    Sample Bar CodesUp to 20 digits; Interleaved 2 of 5; Code 39, Code 128, Codabar; A, B, and special characters (.-+/*$%)
    Reaction CuvettesCapacity of 1000 optical-grade plastic cuvettes
    Reaction BathAir bath, 37°C
    Assay Result CalculationsSMART algorithm software repeats and confirms reactive testing; results available as completed
    Assay TechnologyChemiluminescence using Advanced Acridinium Ester
    Reagent Tray30-position reagent tray refrigerated between 4 and 10°C (39–50°F)
    Reagent Capacity Onboard30 assays
    Reagent ContainersADVIA Centaur® ReadyPack® (various pack sizes available)
    Reagent Integrity ControlBar-code reagent identification, automatic inventory tracking and flagging, calibration valid tracking and flagging, reagent onboard stability tracking and flagging, reagent expired/reagent low flagging
    Onboard Stability28 days (average)
    Reagent MixingADVIA Centaur ReadyPacks automatically rocked onboard
    Ancillary Reagents25-position ancillary reagent tray cooled 4–8°C (39–46°F)
    Auto-calibration/Auto-QCAvailable automatic QC ordering by test, control, date, and time
    QC DataAdvanced QC package includes Levey-Jennings plots and Westgard rules
    Monitor22-inch diagonal high-resolution LCD touchscreen with adjustable height
    Operating SystemMicrosoft® Windows® 7
    System DocumentationOperator manual, quickstart guide, and context-sensitive online help
    Data Storage1,000,000 test/QC results; 500,000 can be historical; can archive to removable media
    Onboard Maintenance LogsYes
    Host InterfaceRS232C bidirectional and host query; Ethernet connection available
    Host QueryYes
    Remote Access & Service1000Base-T Ethernet port for remote access and modem for Siemens Remote Service
    Removable MediaCD, DVD, or USB
    Power Requirements200-240 V at 50/60 Hz; 1.2 kW maximum
    Maximum Water Consumption2.5 liters (0.6 gallons) per hour
    Drain RequirementsMinimum of 2.5 liters (0.6 gallons) per hour
    DimensionsWith monitor: 167 (h) x 196 (w) x 104 (d) cm; 66 (h) x 77 (w) x 41 (d) in. Without monitor: 133 (h) x 196 (w) x 104 (d) cm; 52 (h) x 77 (w) x 41 (d) in.
    Weight585 kg (1289 lb)
    ComplianceComplies with international environmental, health, and safety standards
    Noise Emission65 dB maximum
    Processing Heat Output5100 BTU/hour
    Ambient Temperature18–30°C (64–86°F)
    Ambient Humidity20–80% noncondensing
    Floor Load-Bearing Requirement412 kg/m2 (84 lb/ft2)