Siemens ADVIA Centaur XP Immunoassay System

Siemens ADVIA Centaur XP Immunoassay System

Contact us for price

Description

ADVIA Centaur XP system has more onboard reagents and dedicated STAT capabilities to maximize productivity, regardless of volume or types of tests. Always ready , continuous operation without interruption. The advia Centaur XP system has a High throughput up to 240 tests/hour,Intuitive software for ease of operation, STAT port guarantees priority sampling at any time.

Overview 

  • System Description Random access immunoassay system
  • Throughput- Up to 240 tests per hour
  • Time To First Result 18 minutes; results every 15 seconds thereafter, assay dependent
  • Assays On Board 30
  • Assay Packs ADVIA Centaur® ReadyPack™ assays
  • Assay Technology Direct Chemiluminescence using Acridinium Ester technology
  • Sample Handling
  • Sample Tubes 1 mL, 2 mL, 3 mL, 5 mL, 7 mL, 10 mL, micro container tubes
  • Sample Carriers Universal 5-position rack holds multiple tube types; 180 samples fully loaded; no-pause loading and unloading
  • STAT Handling Dedicated STAT port accepts samples any time
  • Sample Types Serum, plasma, urine – assay dependent
  • Sample Volume 10 uL to 200 uL assay dependent
  • Sample Integrity Control Pressure transducer technology with clot management, detection, and flagging.
  • Sample Pipette Tips Ensure zero sample carryover
  • Automatic Dilutions Varies by assay, up to 1:1000
  • Test Menu

    Reproductive Endocrinology Therapeutic Drug Monitoring Thyroid
    AFP Carbamazepine Anti-TG
    Anti-Müllerian Hormone* Digitoxin Anti-TPO
    DHEAS Digoxin Free T3
    Enhanced Estradiol Gentamicin Free T4
    Free βhCG‡‡ Phenobarbital T Uptake
    FSH Phenytoin Total T3
    hCG Theophylline Total T4
    LH Valproic Acid TSH3-Ultra
    PAPP-A‡‡ Vancomycin TSH
    PIGF*
    Progesterone Immunosuppressant Drugs Bone Metabolism
    Prolactin Cyclosporine Intact PTH
    sFLT-1* Everolimus* Vitamin D Total
    SHBG Tacrolimus*
    Testosterone
    Hepatitis Oncology ToRCH
    Anti-HBs-2 AFP CMV IgG
    HAV IgM BR 27.29 CMV IgM*
    HAV Total CA 125II Herpes I
    HBc IgM CA 15-3 Herpes II
    HBc Total CA 19-9 Rubella IgG
    HBeAg Calcitonin Rubella IgM
    HBsAg Confirmatory CEA Toxoplasma IgG
    HBs AgII Complexed PSA Toxoplasma IgM
    HCV** PSA
    Serum HER-2/neu
    Anemia Cardiac Special ID
    Active B12 BNP EBV-EBNA IgG*
    Ferritin CKMB EBV-VCA IgG*
    Folate High-Sensitivity Troponin I (TNIH) EBV-VCA IgM*
    RBC Folate Myoglobin Syphilis
    Vitamin B12 Tnl-Ultra Zika IgM††
    Metabolic HIV Diabetes
    Cortisol HIV 1/0/2 Enhanced (EHIV)§ C-Peptide
    Homocysteine HIV Combo§ Insulin
    Liver Disease – NASH Autoimmune Inflammation
    ELF™ [Enhanced Liver Fibrosis] Anti-CCP IgE, Total

    Specifications

    ManufacturerSiemens
    ConditionNew
    System DescriptionRandom access immunoassay system
    ThroughputUp to 240 tests per hour
    Time To First Result18 minutes; results every 15 seconds thereafter, assay dependent
    Assays On Board30
    Assay PacksADVIA Centaur® ReadyPack™ assays
    Assay TechnologyDirect Chemiluminescence using Acridinium Ester technology
    Sample Tubes1 mL, 2 mL, 3 mL, 5 mL, 7 mL, 10 mL, micro container tubes
    Sample CarriersUniversal 5-position rack holds multiple tube types; 180 samples fully loaded; no-pause loading and unloading
    STAT HandlingDedicated STAT port accepts samples any time
    Sample TypesSerum, plasma, urine – assay dependent
    Sample Volume10 uL to 200 uL assay dependent
    Sample Integrity ControlPressure transducer technology with clot management, detection, and flagging.
    Sample Pipette TipsEnsure zero sample carryover
    Automatic DilutionsVaries by assay, up to 1:1000
    Reagent Cooling30-position reagent tray cooled 4°C to 8°C
    Reagent Integrity ControlBarcode reagent identification, automatic inventory tracking and flagging, calibration validity tracked and flagged, reagent on board stability tracked and flagged, reagent expired/ reagent low flagging
    Reagent MixingReadyPacks automatically rocked onboard
    Ancillary Reagents25-position ancillary reagent tray cooled 4°C to 8°C
    Cuvettes1,000 ADVIA Centaur cuvette capacity
    Sample TipsADVIA Centaur type, 840 on board, automatic tracking, and flagging of inventory
    Monitor19-inch diagonal high-resolution LCD touch screen with adjustable height
    Operating System Results ManagementDual Sun Microsystems Sparc® II processors; Microsoft® Windows® XP
    Host InterfaceRS232C bidirectional and host query; Ethernet connection available; 25,000 result storage; archive to removable media; automatic reflex testing
    Smart Algorithm SoftwareRepeats and confirms reactive testing; results available as completed
    Electrical208-240 V at 50/60 Hz; meets regulatory requirements for UL, CAN/CSA, IEC1010. JIS, CE, CISPR11 Class A
    Power Consumption1 KW in ready state; 1.5 KW in operating state
    WaterDeionized type II
    Dimensions51.5 (h) x 72.4 (w) x 41.0 (d) in-130.9 (h) x 183.9 (w) x 104.2 (d) cm
    Weight1,200 lbs (545 kg)
    Floor Load-bearing Requirement390 kg/m2 (80lb/ft2)
    Operating Environment Temperature18-30OC; humidity 20-80% noncondensing
    Ventilation10,000 BTU/hour
    Noise Emission58.8 db in ready state; 61.3 in operating state