Roche Cobas E411 Fully Automated Immunoassay Analyzer

Roche Cobas E411 Fully Automated Immunoassay Analyzer

Contact us for price

Description

Roche Cobas E411 Immunoassay Analyzer

 Includes:  

One cobas E411 analyzer with computer, software, bench, and manual.

Features: 

The cobas e 411 analyzer is a fully automated analyzer that uses a patented ElectroChemiLuminescence (ECL) technology for immunoassay analysis. It is designed for both quantitative and qualitative in vitro assay determinations for a broad range of applications (including anemia; bone, cardiac and tumor markers; critical care; fertility/hormones; and infectious diseases). The analyzer is available as a rack or disk sample handling system.

Features/Benefits

  • High sensitivity, reliability, and reproducibility of results due to ECL technology
  • More than 100 assays available1
  • Up to 18 assays can be run at a throughput of up to 86 tests/h2
  • 9 min STAT applications for emergency sample assays (incl. TnT Gen. 5, Creatine-Kinase MB, Myoglobin, Parathyroid Hormone (PTH), and human Chorionic Gonadotropin (hCG)1
  •  

    Test Menu

    Elecsys® ACTH Elecsys® Anti-TPO Elecsys® Cyclosporine Elecsys HBeAg
    Elecsys® AFP Elecsys® Anti-TSHR Elecsys® CYFRA 21-1 Elecsy HBsAg Confirmatory Test
    Elecsys® AMH Elecsys® C-Peptide Elecsys® DHEA-S Elecsys HBsAg II
    Elecsys® Anti-CCP Elecsys® CA 125 II Elecsys® Digoxin Elecsys CG STAT
    Elecsys® Anti-HAV Elecsys® CA 15-3 II Elecsys® Estradiol III Elecsys CG+
    Elecsys® Anti-HAV IgM Elecsys® CA 19-9 Elecsys® Ferritin Elecsys HE4
    Elecsys® Anti-HBc IgM Elecsys® Calcitonin Elecsys® Folate III Elecsys hGH
    Elecsys® Anti-HBc II Elecsys® CEA Elecsys® Folate RBC Elecsys HSV-1 IgG
    Elecsys® Anti-HBe Elecsys® CK-MB Elecsys® free PSA Elecsys HSV-2 IgG
    Elecsys® Anti-HCV II Elecsys® CMV IgG Elecsys® free βhCG Elecsys IgE II
    Elecsys® Anti-SARS-CoV-2 S Elecsys® CMV IgM Elecsys® FSH Elecsys IGF-1
    Elecsys® Anti-Tg Elecsys® Cortisol II Elecsys® FT3 III Elecsys IGFBP-3
    Elecsys® IL-6 Elecsys Prolactin II Elecsys® Tacrolimus
    Elecsys® Insulin Elecsys PTH Elecsys® Testosterone II
    Elecsys® LH Elecsys Rubella IgG Elecsys® total PSA
    Elecsys® Myoglobin Elecsys Rubella IgM Elecsys® Total-Tau CSF
    Elecsys® N-MID Osteocalcin Elecsys S100 Elecsys® Toxo IgG
    Elecsys® NSE Elecsys SARS-CoV-2 Antigen Elecsys® Toxo IgM
    Elecsys® PAPP-A Elecsys sFlt-1 Elecsys® Troponin I
    Elecsys® Phospho-Tau (181P) CSF Elecsys SHBG Elecsys® TSH
    Elecsys® PlGF Elecsys Syphilis Elecsys® Vitamin B12 II
    Elecsys® proBNP II Elecsys T-Uptake Elecsys® Vitamin D total II
    Elecsys® Progesterone III Elecsys T3 Elecsys® β-CrossLaps/serum
    Elecsys® ProGRP Elecsys T4

    Specifications

    ManufacturerRoche Diagnostics
    ConditionNew
    WeightDisk: 180 kg / 397 lbs  Rack: 250 kg / 551 lbs
    SystemFully automated, immunoassay analyzer for random access processing of ECL-based immunoassays (cobas e system format)
    System componentsSelf contained bench top analyzer comprising an analytical unit and a customized user interface
    Types of modulescobas e 411 disk analyzer  cobas e 411 rack analyzer Optional System Table (cabinet); Optional System Table Extension (for printer)
    System interfacesRS232 serial interface, bidirectional
    ThroughputUp to 86 tests/hour
    Number of reagent channels25 reagent slots with a maximum of 18 on-board assays
    Programmable test parametersN/A, programming-by-loading (PBT) concept, application data are transferred without operator intervention from the 2D barcode of the reagent pack (RP) into the instrument database
    Sample materialsSerum/plasma, urine, others
    Sample input/output30 positions for samples, calibrators and controls  15 racks with 5 samples each (= 75 samples in/out)
    Sample volume10 – 50 μl.
    Sample clot detectionStandard (pressure sensor)
    Sample barcode typesCode 128; Codabar (NW 7); Interleaved 2 of 5; Code 39
    Control unitMicrosoft® Windows® XP-based panel PC
    Calibrator/QC inputcobas e system-specific barcoded CalSet vials on disk or racks
    Calibration methodsLot calibration (L-cal); Reagent Pack (RP) calibration (R-Cal)
    QC methodsIndividual QC and cumulative QC  Up to 100 controls pre-programmable
    Data storage capacityThe memory contains data files necessary for the analyzer and software to work together:  Reagent Data File: Up to 300 reagent packs  Sample Data File: Up to 2000 test records (for samples and controls)  Calibration Data File: Up to 160 calibrators – QC Data File: Capacity up to 100 controls  Operating Parameter Data File: Up to 305 reagent applications  Up to 20 operator IDs
    Electrical requirements230/110 Volts AC; 1,000 kVA (disk), 1,250 kVA (rack)  Frequency: 50 Hz or 60 Hz +/- 0.5%
    Water/Waste requirementsWater: Bacteria free, de-ionized water supply, resistance of less than 10 μS/cm  Liquid waste: Onboard waste container (4 liter), direct drain optional
    Operating conditionsAmbient temperature: 18 to 32°C  Ambient humidity: 20 to 80% RH (without condensation)  Heat Output: 2,879 kJ/hr (analyzer unit)  Noise Output: 60 dBA (stand-by), 63 dBA (operation avg.)
    Physical dimensionsWidth (disk/rack): 120 cm / 170 cm 47.2 in / 67 in  Depth (disk/rack): 73 cm / 95 cm 28.7 in / 37.4 in  Height: 80 cm / 31.4 in (closed top cover) 109 cm / 43 in (opened top cover)