QIAGEN QIAxpert Spectrophotometer

QIAGEN QIAxpert Spectrophotometer

Contact us for price

Description

If you are looking for a smarter solution for assessing nucleic acid quantity and quality, look no further. The QIAxpert spectrophotometer provides a detailed insight into your sample’s quantity and purity by analyzing your nucleic acid with spectral content profiling. This proprietary analysis feature lets you specifically quantify DNA, RNA and total nucleic acids and also provides information regarding residual impurities. The QIAxpert helps you understand your samples better than ever before. Be aware of work references real money slots online. In addition to classical UV-Vis quantification of nucleic acids, QIAxpert can differentiate DNA, RNA and sample impurities, giving you a clear picture of what really is in your samples.
Along with the next-generation analysis protocols, QIAxpert users particularly like the clean and simple way of loading samples, the fast measurements and the comprehensive reports that can be automatically exported. QIAxpert features a full-color integrated touchscreen interface, so no extra computer is required. The intuitive and easy-to-use QIAxpert Software was designed to fit a busy lab schedule. To facilitate data analysis and sharing, the data are exportable to a USB stick, a networked drive, or a smart device via a QR code.

Features

  • Quantify nucleic acids from up to 16 samples in less than 2 minutes
  • Discriminate between molecules of interest using unique spectral protocols
  • Determine specific amounts of DNA, RNA,and other contaminating fractions
  • Rapidly perform analyses via the intuitive, full-color integrated touchscreen
  • Generate comprehensive reports to view on any computer or smart device
  • QIAxpert is an innovative high-speed microfluidic UV/VIS spectrophotometer which profiles sample content to differentiate between DNA and RNA and sample impurities. QIAxpert is fast and easy to use, analyzing up to 16 samples within 2 minutes — just pipette samples onto the QIAxpert slide, place it into the reader, and select the pre-programmed method on the integrated touchscreen. QIAxpert is the right tool for any fast-paced lab performing nucleic acid quantification and quality control.

    ~High system linearity for best results

    ~Proven reproducibility of measurements

    ~Flexible report options to meet your needs.

    ~Flexibility and accuracy with QIAxpert Slides.

    QIAxpert uses disposable microfluidic slides to analyze up to 16 samples per run. On-board capillary intake and storage channels ensure easy sample loading with no cross-contamination, and allow unused positions to be used in later analyses. The capillary channels preserve samples and prevent evaporation. Thus, samples can be loaded and left at room temperature for up to 2 hours prior to readout. Comprehensive reports of measurements are available in several formats for data analysis and sharing.

    Only 2 µl of sample required Easy sample loading with capillary intake and storage Samples are stable for up to 2 h with no evaporation Flexible loading of 1–16 samples Unused positions on slide can be used later No cross-contamination, no cleaning step 0.5 mm optical path; 0.03 to 40 OD (10 mm equivalent) A 260)

    Specifications

    ManufacturerQiagen
    ConditionNew
    FeaturesSpecifications
    Absorbance precision0.001 OD
    Absorbance accuracy4% (0.66 OD at 350 nm)
    Concentration range1.5 ng/μl up to 2000 ng/μl dsDNA (A260 dsDNA application)
    Data outputUSB, TCP/IP or QR code
    Dimensions(W) 23 x (D) 28 x (H) 30 cm; 9 kg
    Light sourceXenon flash lamp
    Maximum sample residence time2 hours
    Measurement range (10 mm equivalent)0.03–40 OD
    Measurement timeApproximately 2 minutes for fully loaded slide
    Path length0.5 mm
    Optical principleUltrasensitive photodiode array spectrophotometer for UV-VIS range
    Photometric range0.0005–2.0 OD
    Power supply100–240 V, 50/60Hz (operating voltage 24 VDC)
    Recommended sample quantity2 µl
    Samples per run (throughput)1–16 samples per run (blanks included)
    SoftwareOn-board software; newest software can be downloaded at www.qiagen.com/QIAxpert
    TypeUV/VIS polychromatic system + reference channel
    Wavelength accuracy≤0.2 nm
    Wavelength range230–750 nm
    Wavelength reproducibility0.1 nm
    Wavelength resolution<3 nm
    DNA range1.5–2000 ng/μl dsDNA (