Awareness Technology Stat Fax 3300 Chemistry Analyzer

Awareness Technology Stat Fax 3300 Chemistry Analyzer
Contact us for price
Description
The Awareness Technology Stat Fax 3300 Microplate Reader is designed to read and calculate the results of in-vitro diagnostic assays. Equipped with a large graphical display, the reader provides absorbance and concentration readings at or near the available wavelength. The flexible Stat Fax 3300 is equipped with multiple, self-prompting operating modes to meet the varying demands and requirements of any facility. The modes include Absorbance Mode, Standard Mode, Rate Mode, Factor Mode, Multi-point Mode, and Index Mode. Moreover, the instrument accepts standard 12 mm round tubes as well as 1 cm square cuvettes.
Features/Benefits
- Optical system temperature controlled to 37 C
- Bi-chromatic with 6 filters and spaces for 2 optional additional filters (340-700nm)
- Adapts for use with 12mm round tubes, 1cm square cuvettes, or built-in flow cell
- Vacuum system speeds sampling and reduces carryover
- Calibration traceable to NIST standard
- Use the optional external keyboard for easy entry of patient names and data
- Use Redi-Check to monitor accuracy, linearity, and repeatability the easy way
- Peltier cell cooling available as an option
- View kinetic plots in real time
- Calculate kinetic results based on the entire read time OR the segment you choose.
Stat Fax 3300 stores patient data
- Create a work list.
- Run the tests.
- Print Patient Reports.
Output data to:
- On-board 40-column
- Thermal printer
- External printer
- PC and LIS.
Stat Fax 3300 stores control values
- and of fers many QC options to assure linear kinetics and reliable results.
- Create Levey-Jennings plots for QC
Stat Fax 3300 stores calibrations and standard curves
- For each run, choose a new curve or use a stored curve with options to revert to the previous curve and to modify multi- point curves.
Specifications
| Manufacturer | Awareness Technology |
| Condition | New |
| Stock Number | 124 |
| Spectrophotometer Type | Filter photometer |
| Optical Configuration | Single beam with continuously rotating filter wheel Monochromatic or bichromatic reading 8 filter positions |
| Usable Spectral Range | 330 to 770 nm |
| System Procedures | Open and by stored menu |
| Calculating Modes | Absorbance, Single Standard, Factor Mode, Multi Standard Mode (up to 7 standards), Multi Standard % Abs (up to 7 standards), Kinetic Mode, Index Mode |
| Channels | 120 Open |
| Source of Radiation | Tungsten Halogen, 10 Watt, with automatic lamp saver |
| Selection of Wavelength | By filter |
| Filter Type | 4-cavity interference, long-life ion beam-assisted deposition |
| Wavelength Accuracy | +/- 3 nm |
| Filter Location | After sample (heat absorbing filter before sample) |
| Filter Selection | Automatic by software or via keyboard |
| Wavelengths | 340, 405, 505, 545, 580, 630 nm supplied standard, other/additional filters optional |
| Half Bandwidth | < 10 nm |
| 1/100 Bandwidth | 14 nm at 340 nm |
| False Radiant Energy Ratio | < 0.001 at 340 and 405 nm |
| Cuvette | 1 cm square, 12 cm round, flow through |
| Supplied Type | flow through |
| Material | 316 stainless, borosilicate windows |
| Geometry | Cylindrical, 2.3 mm dia x 5 mm +/- 0.05 mm |
| Illuminated Volume | 21 µl |
| Minimum Read Volume | 250 µl |
| Aspiration/Purge | Vacuum pump at 18 cm of Hg |
| Valve | Silicone pinch type |
| Cuvette Holder | Thermostatically controlled compartment at 37° C |
| Detector | Gallium-Arsenide-Phosphide photodiode |
| Display Type | 240 x 128 Graphic LCD w/ Backlight |
| Absorbance | -0.5 to 3.5 (flow-through mode) -0.5 to 2.5 (tube or 1 cm cuvette) |
| Concentration | Maximum 999,999 |
| Kinetic Results | Abs/min with resolution of 0.0002 A/min |
| Zero Compensation | Automatic |
| Range | -0.5 to 2.5 absorbance |
| Parallel | Centronics/IBM-PC compatible |
| Serial | RS-232 at 9600 baud, 8 data, 1 stop, no parity Bi-Directional |
| Data Input | 1) 20 Key Keypad, 2) PS2 101 Keyboard (connector available at back of instrument) |
| Flow-through | < 0.5 % at 1 absorbance, 340/630 nm |
| Stability | Better than 0.003 A/hr monochromatic after warm-up Better than 0.001 A/hr bichromatic after warm-up |
| Warm-up Time | 90 seconds photometric 15 minutes for temperature compartment |
| Electronics | Z180 Microprocessor 18 MHz 128k EEPROM 32 K bytes Nonvolatile RAM (NVRAM) |
| Power | Auto-Switching Power Supply Voltage source: 90 – 264 VAC Frequency: 50/60 Hz Power consumption: 60 watts Installation Category: CAT II Fuses: 2.5A/250V Fast 5-20 mm Glass Fuse, 2.0A/250V Fast 5-20 mm Ceramic Fuse, (2) 6/10 250V Slow Blow 3AG Fuses |
| Dimensions and Weight | 40cm (L) x 37 cm (W) x 14 cm (H) lid closed (30 cm lid opened), 6.4 kg |