Roche Cobas E411 Fully Automated Immunoassay Analyzer

Roche Cobas E411 Fully Automated Immunoassay Analyzer
Contact us for price
Description
Roche Cobas E411 Immunoassay Analyzer
Includes:
One cobas E411 analyzer with computer, software, bench, and manual.
The cobas e 411 analyzer is a fully automated analyzer that uses a patented ElectroChemiLuminescence (ECL) technology for immunoassay analysis. It is designed for both quantitative and qualitative in vitro assay determinations for a broad range of applications, including:
- Anemia
- Bone, cardiac, and tumor markers
- Critical care
- Fertility/hormones
- Infectious diseases
The analyzer is available with either a rack or disk sample handling system.
Features/Benefits:
- High sensitivity, reliability, and reproducibility of results due to ECL technology
- More than 100 assays available¹
- Up to 18 assays can be run at a throughput of up to 86 tests/hour²
- 9 min STAT applications for emergency sample assays, including:
- TnT Gen. 5
- Creatine-Kinase MB
- Myoglobin
- Parathyroid Hormone (PTH)
- Human Chorionic Gonadotropin (hCG)¹
Test Menu
Elecsys® ACTH
Elecsys® Anti-TPO
Elecsys® Cyclosporine
Elecsys HBeAg
Elecsys® AFP
Elecsys® Anti-TSHR
Elecsys® CYFRA 21-1
Elecsy HBsAg Confirmatory Test
Elecsys® AMH
Elecsys® C-Peptide
Elecsys® DHEA-S
Elecsys HBsAg II
Elecsys® Anti-CCP
Elecsys® CA 125 II
Elecsys® Digoxin
Elecsys CG STAT
Elecsys® Anti-HAV
Elecsys® CA 15-3 II
Elecsys® Estradiol III
Elecsys CG+
Elecsys® Anti-HAV IgM
Elecsys® CA 19-9
Elecsys® Ferritin
Elecsys HE4
Elecsys® Anti-HBc IgM
Elecsys® Calcitonin
Elecsys® Folate III
Elecsys hGH
Elecsys® Anti-HBc II
Elecsys® CEA
Elecsys® Folate RBC
Elecsys HSV-1 IgG
Elecsys® Anti-HBe
Elecsys® CK-MB
Elecsys® free PSA
Elecsys HSV-2 IgG
Elecsys® Anti-HCV II
Elecsys® CMV IgG
Elecsys® free βhCG
Elecsys IgE II
Elecsys® Anti-SARS-CoV-2 S
Elecsys® CMV IgM
Elecsys® FSH
Elecsys IGF-1
Elecsys® Anti-Tg
Elecsys® Cortisol II
Elecsys® FT3 III
Elecsys IGFBP-3
Elecsys® IL-6
Elecsys Prolactin II
Elecsys® Tacrolimus
Elecsys® Insulin
Elecsys PTH
Elecsys® Testosterone II
Elecsys® LH
Elecsys Rubella IgG
Elecsys® total PSA
Elecsys® Myoglobin
Elecsys Rubella IgM
Elecsys® Total-Tau CSF
Elecsys® N-MID Osteocalcin
Elecsys S100
Elecsys® Toxo IgG
Elecsys® NSE
Elecsys SARS-CoV-2 Antigen
Elecsys® Toxo IgM
Elecsys® PAPP-A
Elecsys sFlt-1
Elecsys® Troponin I
Elecsys® Phospho-Tau (181P) CSF
Elecsys SHBG
Elecsys® TSH
Elecsys® PlGF
Elecsys Syphilis
Elecsys® Vitamin B12 II
Elecsys® proBNP II
Elecsys T-Uptake
Elecsys® Vitamin D total II
Elecsys® Progesterone III
Elecsys T3
Elecsys® β-CrossLaps/serum
Elecsys® ProGRP
Elecsys T4
Specifications
| Manufacturer | Roche Diagnostics |
| Condition | New |
| Stock Number | 281 |
| Weight | Disk: 180 kg / 397 lbs Rack: 250 kg / 551 lbs |
| System | Fully automated, immunoassay analyzer for random access processing of ECL-based immunoassays (cobas e system format) |
| System components | Self contained bench top analyzer comprising an analytical unit and a customized user interface |
| Types of modules | cobas e 411 disk analyzer cobas e 411 rack analyzer Optional System Table (cabinet); Optional System Table Extension (for printer) |
| System interfaces | RS232 serial interface, bidirectional |
| Throughput | Up to 86 tests/hour |
| Number of reagent channels | 25 reagent slots with a maximum of 18 on-board assays |
| Programmable test parameters | N/A, programming-by-loading (PBT) concept, application data are transferred without operator intervention from the 2D barcode of the reagent pack (RP) into the instrument database |
| Sample materials | Serum/plasma, urine, others |
| Sample input/output | 30 positions for samples, calibrators and controls 15 racks with 5 samples each (= 75 samples in/out) |
| Sample volume | 10 – 50 μl. |
| Sample clot detection | Standard (pressure sensor) |
| Sample barcode types | Code 128; Codabar (NW 7); Interleaved 2 of 5; Code 39 |
| Control unit | Microsoft® Windows® XP-based panel PC |
| Calibrator/QC input | cobas e system-specific barcoded CalSet vials on disk or racks |
| Calibration methods | Lot calibration (L-cal); Reagent Pack (RP) calibration (R-Cal) |
| QC methods | Individual QC and cumulative QC Up to 100 controls pre-programmable |
| Data storage capacity | The memory contains data files necessary for the analyzer and software to work together: Reagent Data File: Up to 300 reagent packs Sample Data File: Up to 2000 test records (for samples and controls) Calibration Data File: Up to 160 calibrators – QC Data File: Capacity up to 100 controls Operating Parameter Data File: Up to 305 reagent applications Up to 20 operator IDs |
| Electrical requirements | 230/110 Volts AC; 1,000 kVA (disk), 1,250 kVA (rack) Frequency: 50 Hz or 60 Hz +/- 0.5% |
| Water/Waste requirements | Water: Bacteria free, de-ionized water supply, resistance of less than 10 μS/cm Liquid waste: Onboard waste container (4 liter), direct drain optional |
| Operating conditions | Ambient temperature: 18 to 32°C Ambient humidity: 20 to 80% RH (without condensation) Heat Output: 2,879 kJ/hr (analyzer unit) Noise Output: 60 dBA (stand-by), 63 dBA (operation avg.) |
| Physical dimensions | Width (disk/rack): 120 cm / 170 cm 47.2 in / 67 in Depth (disk/rack): 73 cm / 95 cm 28.7 in / 37.4 in Height: 80 cm / 31.4 in (closed top cover) 109 cm / 43 in (opened top cover) |