Roche Cobas 6000 (c501 + e601) Chemistry Analyzer

Roche Cobas 6000 (c501 + e601) Chemistry Analyzer
Contact us for price
Description
The Cobas 6000 analyzer series offers tailor-made solutions for medium to large workload laboratories. Flexible combinations of clinical chemistry (Cobas C501 module) and immunochemistry testing (Cobas E601 module) modules are available.
Core Unit
- Simple operation with continuous loading and unloading
- Loading capacity of 150 samples in two trays of 75
- Load 5-position racks via dedicated STAT port with rerun buffer or by tray
- Unique 5-position rack for more efficient sample handling
- Smart track rotor for intelligent sample handling
- Single control unit for all modules
Cobas C501 Module
- Clinical chemistry, homogeneous immunoassays, ISE, HbA1c (whole blood measurement)
- Throughput of up to 1,000 tests/hour
- 60 reagent channels plus 3 ISEs
- Reagent loading on the fly
- Specimen integrity via serum indices (parallel processing), clot and liquid level detection
- Contact-free ultrasonic mixing
- Automated rerun and dilution
Cobas E601 Module
- Heterogeneous immunoassays including cardiac markers, anemia, bone and tumor markers, hormones, and infectious diseases
- Electrochemiluminescence (ECL) technology
- Throughput up to 170 tests/hour
- 25 reagent channels
- Carry-over free disposable tips and cups
- 9-minute STAT applications for Troponin I and T, proBNP II, Myoglobin, hCG, and CKMB
- Clot and liquid level detection
- Automated rerun and dilution
Escape from Corporate Service Contracts in California
The quality of service contracts from manufacturers has declined over the last 10 years. They are less responsive, slower, and more corporatized. With IDE:
- Your service engineer is local in California
- Call or text your service engineer directly
- Same day phone service, next-day on-site service in California, next-flight service anywhere in the US
- Faster repairs because parts are kept in stock
Increase Throughput, Uptime, and Value of Every Test
- Your refurbished Roche Cobas 6000 is returned to like-new specs
- Tens of thousands of dollars less than new systems
- Investment pays for itself within 2 years through efficiency and reduced downtime
- Financing options available
- Extend the lifespan of your primary analyzers by 5+ years
Make Onboarding New Staff Easier in California
- Refurbished systems reduce training time for inexperienced staff
- Error rates drastically reduced
- Oversight time reduced in the first 9 months
- Easy adoption because many candidates are already familiar with Cobas systems
Get Personal, Responsive Service in California
- Direct access to your service engineer
- Family-style business approach
- Guidance provided until system is fully operational
- Avoid frustration with large, centralized service providers
All-in-One Support for Your Roche Cobas System
- Parts, reagents, and preventive maintenance in one place
- Avoid cobbling together multiple suppliers
- Trade-in options for old analyzers to save on upgrades
- Maintain low environmental impact
Service Options in California
- Preventive maintenance to reduce downtime
- Tailored service agreements
- Same-day phone support
- Affordable depot service for repairs or refurbishment
- Decontamination and relocation assistance
- Installation, IQ/OQ validation, calibration, quality control, and professional training
COBAS C501 Module
6AM
A1C-3
A1MG2
A2MG
AAGP2
AAT2
ACET2
ALB2
ALBP
ALP2
ALT
ALTL
ALTLP
AMIK2
AMPS2
AMY-P
AMYL2
APOAT
APOBT
ASLOT
ASTL
ASTLP
B2MG
BARB
BENZ
BILT3
BUP
C3C-2
C4-2
CA2
CARB4
CERU
CHE
CHE2
CHOL2
CK
CO2-L
COC2
CREJ2
CREP2
CRP4
CRPHS
CRPLX
CYSC2
D Bili
D-DI
DIG
EDDP
ETOH2
FEN2
FERR4
FRA
GENTC
GGT-2
GLUC2
GLUC3
HAPT2
HCYS
HYD
IGA-2
IGG-2
IGM-2
LACT2
LDHI2
LDLC3
LFLC
LI
LPA2
LSD
MDN2
Methadone Metabolite
MG2
MPA
MTQL
MYO2
NAPA2
NH3L
NH3L2
OPI2
OXY
PCP
PHNO2
PHNY2
PPX
PREA
PROC2
QUIN2
RF-II
SALI
SI2
STFR
SVTCH
SVTCR
SVTNI
SVTOX
SVTpH
SVTSG
THC2
THEO2
TOBR2
TP2
TPUC3
TRIGL
UA2
UIBC
UREAL
VALP2
VANC3
β2-Microglobulin
COBAS E601 Module
Elecsys ACTH
Elecsys AFP
Elecsys AMH
Elecsys Anti-CCP
Elecsys Anti-HAV
Elecsys Anti-HAV IgM
Elecsys Anti-HBc IgM
Elecsys Anti-HBc II
Elecsys Anti-HBe
Elecsys Anti-HCV II
Elecsys Anti-SARS-CoV-2 S
Elecsys Anti-Tg
Elecsys Anti-TPO
Elecsys Anti-TSHR
Elecsys BRAHMS PCT
Elecsys C-Peptide
Elecsys CA 125 II
Elecsys CA 15-3 II
Elecsys CA 19-9
Elecsys Calcitonin
Elecsys CEA
Elecsys CK-MB
Elecsys CMV IgG
Elecsys CMV IgM
Elecsys Cortisol II
Elecsys Cyclosporine
Elecsys CYFRA 21-1
Elecsys DHEA-S
Elecsys Digoxin
Elecsys Estradiol III
Elecsys Ferritin
Elecsys Folate III
Elecsys Folate RBC
Elecsys free PSA
Elecsys free βhCG
Elecsys FSH
Elecsys FT3 III
Elecsys HBeAg
Elecsys HBsAg Confirmatory Test
Elecsys HBsAg II
Elecsys HCG STAT
Elecsys HCG+β
Elecsys HE4
Elecsys hGH
Elecsys HSV-1 IgG
Elecsys HSV-2 IgG
Elecsys IgE II
Elecsys IGF-1
Elecsys IGFBP-3
Elecsys Insulin
Elecsys LH
Elecsys Myoglobin
Elecsys NSE
Elecsys N-MID Osteocalcin
Elecsys PAPP-A
Elecsys Phospho-Tau (181P) CSF
Elecsys PlGF
Elecsys proBNP II
Elecsys Progesterone III
Elecsys ProGRP
Elecsys Prolactin II
Elecsys PTH
Elecsys Rubella IgG
Elecsys Rubella IgM
Elecsys S100
Elecsys SARS-CoV-2 Antigen
Elecsys sFlt-1
Elecsys SHBG
Elecsys Syphilis
Elecsys T-Uptake
Elecsys T3
Elecsys T4
Elecsys Tacrolimus
Elecsys Testosterone II
Elecsys Total-Tau CSF
Specifications
| Manufacturer | Roche Diagnostics |
| Condition | New |
| Stock Number | 283 |
| System | 2nd Generation SWA: modular, analytical system platform, consolidated work area for Clinical Chemistry and Immunology, expandable and re-configurable on site |
| Types of modules | · cobas c 501 module: Photometric measuring unit (incl. ISE) · cobas e 601 module: Electrochemiluminescence (ECL) technology measuring unit |
| Number of configurations | 7 module combinations; Up to 3 modules in one core unit · cobas c 501 modules: 1 to 2 modules · cobas e 601 modules: 1 to 2 modules · cobas c 501 | cobas e 601 modules: 2 to 3 modules (only 2 units of 1 module type for SWA combinations possbile) |
| Sample throughput | Up to 120 racks/hour or 600 samples/hour |
| Test throughput (theoretical max) | · 170 up to 2,170 tests/hour: overall · 1,000 up to 2,000 tests/hour:with cobas c 501 modules · 170 up to 340 tests/hour:with cobas e 601 modules · 1,170 up to 2,170 tests/hour: with cobas c 501 and cobas e 601 modules |
| Number of reagent channels (reagent slots) | Up to 151 channels, in total 63 channels (including ISE, automatic setting) on cobas c 501 module 25 channels on cobas e 601 module |
| Sample material | Serum, Plasma, Urine, CSF |
| Core unit analytics | · Rack: 5 position rack, RD standard rack · Rack types: Routine, STAT, Control, Calibrator, Rerun (manual), Wash · Tray: Tray with 15 racks/75 samples, RD standard tray · STAT port: STAT samples are processed with true STAT interrupt |
| Sample container types | · Primary tubes: 5 to 10 mL; 16 x 100, 16 x 75, 13 x 100, 13 x 75 mm · Sample cup: 2.5 mL · Micro cup: 1.5 mL, (exception: cobas e 601 module) · Cups on tube: Cup on top of a 16 x 75/100 mm tube; Cup on top of one non standard tube · False bottom tube (FBT): One type is definable |
| Sample volume | 1–35 µL |
| Sample dilution | 3–121 times |
| Sample clot detection | Available for cobas c 501 and cobas e 601 module |
| Minimum sample volume | · Primary tubes: 700 µL · Sample cup: 100 µL · Micro cup (FTB): 50 µL |
| Sample barcode types | Code 128; Codabar (NW 7); Interleaved 2 of 5; Code 39 |
| System interfaces | RS 232 serial interface, bi-directional Interface to cobas link data station for cobas teleservice functionalities and automatic download |
| Sample data base | 10.000 routine/STAT samples |
| Test methods | For photometric modules: 1 point, 1 point + prozone check, 2 point, 2 point kinetic, 2 point + prozone check, 3 point, 1 point + kinetics Rate A, Rate A + serum index, Rate A with blank, Rate B |
| Calibrator/QC input | Via specified racks through the input buffer or STAT port |
| Calibration methods | Start-up, Re-calibration For photometric modules: Linear, non-linear multi-points, 2 point calibration, K-factor up to 100 different calibrators pre-programmable Storage of up to 180 curves Preventive calibration of the stand-by cobas c and cobas e packs on cobas c 501: Two k-factor can be defined for different sample types |
| QC methods | For e- and photometric modules: Real-time QC, individual QC, cumulative QC Up to 100 controls pre-programmable Preventive QC after calibration of the stand-by cassettes/rack packs Auto QC: QC without operator intervention (timer triggered) |
| Rerun/reflex function | Automatic rerun and manual rerun; Automatic reflex is supported by the system, reflex request to be provided by Roche Middleware Solutions |
| Electrical requirements | · Power requirements: 230 Volts AC; 110 Volts AC 2.4 kVA (for cobas c 501 | e 601) · Frequency: 50 Hz or 60 Hz +/- 0.5% |
| Water/waste requirements | · Water: Bacteria free, deionised water supply: resistance of < 1,0 µS/cm · Water pressure: 0.5 – 3.5 kg/cm3, (49–343 Kpa) · Biohazards waste: Separate container behind the system · L. concentrated: Central drain port, diameter: = 50 mm / 2 inches · Waste drain < waste: 100 mm above the floor / 4 inches |
| Water consumption | · Max flow rate: cobas c 501 module: approx. 40 L/hour · Typical consumption: cobas e 601 module: approx. 30 L/hour |
| Operational conditions | · Ambient temperature: 15 to 32°C / 59 to 90°F · Ambient humidity: 30 to 85% (RH, without condensation) · Heat output: 1.5 kW (5400 kJ/hr for cobas c 501, 4320 kJ/hr for cobas e 601) · Noise output: < 65 dB |
| Physical dimensions | · Width: 188 to 498 cm / 74 to 196 inches · Core unit: 69 cm / 27.2 inches cobas c or cobas e module: 120 cm / 47.2 inches · Depth: cobas c or cobas e module: 98 cm / 38.6 inches Core unit: 104 cm / 41 inches · Height: 130 cm / 51 inches · Control unit: Width: 80 cm / 31.5 inches, depth: 80 cm / 31.5 inches |
| Weight | · cobas c 501 module: 330 kg / 730 lbs · cobas e 601 module: 360 kg / 795 lbs · Core unit (sample loader, control unit, rack rotor): 180 kg / 400 lbs |